Interpretation of in-air output ratio of wedged fields in different measurement conditions


1 Department of Medical Physics, School of Medicine, Tabriz University of Medical Sciences, Tabriz, Iran
2 Department of Medical Physics, School of Paramedical, Kerman University of Medical Sciences, Kerman, Iran

Correspondence Address:
Dr. Ali Jomehzadeh
Department of Medical Physics, School of Paramedical, Kerman University of Medical Sciences, Kerman
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmss.JMSS_36_18

Rights and Permissions

Background: The collimator scatter factor (Sc) is one of the most important parameters in monitor unit (MU) calculation. There are several factors that impact Scvalues, including head structures, backscatter in dose monitoring chambers, and wedges. The objective of this study was to investigate the variation of Scwith different buildup cap materials, wall thickness of buildup caps, source-to-chamber distances (SCDs), ionization chambers, and wedge angles in 6 MV photon beam. Methods: In this study, copper and Perspex buildup caps were made with two different thicknesses for each buildup cap. Measurements were performed on an Elekta Compact medical linear accelerator (6 MV) using RK dosimeter with a sensitive volume of 0.120 cm3 and Farmer-type ion chamber with a sensitive volume of 0.65 cm3. In all measurements, buildup caps and ionization chambers were positioned such as to stand vertically to the beam central axis. It was also investigated the effect of internal wedge with different angles (30° and 60°) different SCDs on Sc. Results: It was found in large field sizes, Scvalues in Perspex buildup cap were higher than copper. Different SCDs and type of ion chamber and wall thickness of buildup caps had no significant influence on Scvalues. The presence of wedge influenced Scvalues significantly. Variation of Scin wedged fields compared to open fields had a maximum deviation of 0.9% and 6.8% in 30° and 60° wedge angles, respectively. Conclusion: It was found that the presence of wedges had a significant influence on Scand increases with wedge angles. As such, it should be taken into account in manual MU calculations.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed78    
    Printed18    
    Emailed0    
    PDF Downloaded16    
    Comments [Add]    

Recommend this journal