ORIGINAL ARTICLE
Year : 2020  |  Volume : 10  |  Issue : 1  |  Page : 48-52

Evaluation of the lung dose in three-dimensional conformal radiation therapy of left-sided breast cancer: A phantom study


1 Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran
2 Department of Radiotherapy Oncology, Seyed Alshohada Hospital, Isfahan University of Medical Sciences, Isfahan, Iran

Date of Submission01-Jan-2019
Date of Decision07-May-2019
Date of Acceptance30-May-2019
Date of Web Publication06-Feb-2020

Correspondence Address:
Prof. Mohammad Bagher Tavakoli
Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmss.JMSS_1_19

Rights and Permissions
  Abstract 


Background: Three-dimensional 3D-CRT: conformal radiation therapy is a selective modality in many radiotherapy centers for the treatment of breast cancer. One of the most common side effects of this method is radiation lung injury. Considering such an injury, lung dose deserves to be studied in depth. Methods: Computed tomography scan of a node-positive left-sided breast cancer woman was used for generating a thorax phantom. Ten thermoluminescent dosimeters (TLDs) were distributed evenly in the left lung of the phantom, and the phantom was scanned. The optimal plan, including supraclavicular and tangential fields, was created by the treatment planning system (TPS). The results of TLD dose measurements at the selected points in the phantom were compared to TPS dose calculations. Results: Lung doses calculated by TPS are significantly different from those measured by the TLDs (P = 0.007). The minimum and maximum differences were −0.91% and 4.46%, respectively. TLDs that were on the inner margin of the lung and breast tissue showed higher dose differences than the TLDs in the lung. Conclusion: The results of this study showed that TPS generally overestimated doses compared to TLD measurements due to incorrect beam modeling caused by contaminated electrons in the lung.

Keywords: Left-sided breast cancer, lung dose, phantom, three-dimensional conformal radiation therapy


How to cite this article:
Abdemanafi M, Tavakoli MB, Akhavan A, Abedi I. Evaluation of the lung dose in three-dimensional conformal radiation therapy of left-sided breast cancer: A phantom study. J Med Signals Sens 2020;10:48-52

How to cite this URL:
Abdemanafi M, Tavakoli MB, Akhavan A, Abedi I. Evaluation of the lung dose in three-dimensional conformal radiation therapy of left-sided breast cancer: A phantom study. J Med Signals Sens [serial online] 2020 [cited 2020 Feb 24];10:48-52. Available from: http://www.jmssjournal.net/text.asp?2020/10/1/48/277817




  Introduction Top


Breast cancer is one of the most common types of cancer in women worldwide.[1] It is often diagnosed at an early stage and treated with surgery, radiotherapy (RT), and systemic therapy.[2],[3] Over the past decades, treatment modalities have evolved to deliver the highest dose to the tumor, while minimizing the radiation dose to normal tissues.[4] Recently, three-dimensional conformal radiotherapy (3D-CRT) and intensity-modulated RT (IMRT) have been adopted in RT centers for the treatment of breast cancer.[5]

In breast cancer RT, 3D-CRT is superior to IMRT because it uses less complicated techniques and minimizes the heart, lung, and contralateral breast dose.[6],[7]

Lung dose assessment in breast cancer patients is challenging because of the deformities of the breast or chest wall[8] and the dose-dependent side-effects.[9],[10] The importance of such an assessment increases in high-risk breast cancer women with lymph node involvement, where tangential and supraclavicular fields are merged and a large part of the lung is within the radiation volume.[11]

Thermoluminescent dosimetry (TLD) measurement is a reliable method for the verification of many dosimetric aspects of an external beam RT.[12] It is the gold standard dosimetry program recommended for quality assurance of machine calibration, planning dosimetry, and dose calculation.[13]

Several studies evaluated lung dose in cancer patients treated with RT. Butson et al.[14] calculated lung dose in an anthropomorphic phantom irradiated with the anterior–posterior field and compared treatment planning system (TPS) dose calculations with TLD dose measurements. Their results showed a 5% TPS dose-overestimation compared to TLD dose measurements. Similar results were found in a study by Davidson et al.[15] They compared lung dose in an IMRT plan with values measured by TLD chips placed in an anthropomorphic phantom. The results of their study showed that TPS overestimated dose levels 10%–15% compared to TLD dose measurements.

The aim of this study was to assess lung dose in left-sided breast cancer treated with 3D-CRT. The results of TLD dose measurements were compared at the same selected points in a thorax phantom with TPS dose calculations.


  Methods Top


Study design

The 3D model of the phantom was generated using computed tomography (CT) images of a left-sided breast cancer patient with 3D-DOCTOR software [Figure 1]. According to ICRU Report Number 44,[16] polymethyl methacrylate with a mass density of 1.18 g/cm3 has been used as normal tissue. The Cork, Plexi, and Teflon were used as lung, heart and spine simulating materials, respectively.
Figure 1: Phantom picture during breast radiotherapy

Click here to view


The phantom has several TLD-positioning holes at various locations [Figure 2]. The holes were numbered to enable dose assessment in the exact positions for each measurement. The dose was measured at different depths of the phantom, which would not be possible in in vivo dosimetry.
Figure 2: TLD positions on a computed tomography scan of a phantom

Click here to view


The phantom used in this study did not have any attachments as breasts because we only intended to compare two different measurement techniques.

TLD calibration

TLD chips used in this study were LiF (TLD-100, NE Technology) with a cross-section of 3 mm × 3 mm and a thickness of 0.9 mm. They are widely used because of their small size, their independent energy response in our study range (4 to10 MV), and their ability to provide high spatial resolution with acceptable precision.[17] They also determine actual doses administered at either skin or body cavities of patients undergoing RT.[18]

Initially, the TLDs were annealed in a TLD oven at 400°C for 1 h and then at 100°C for 2 h. For calibration, all TLDs were placed on a 1-mm Perspex slab and distributed within a 10 cm × 10 cm field at 100 cm source to surface distance (SSD), and a 5-cm slab was placed on top of the Perspex slab to create a build-up region. The 6-MV photon beam was delivered for TLD irradiation at a dose rate of 200 MU/min. The individual correction factor was then calculated by Eq. 1.



<TLD> and TLD are the average reading of the total TLDs and individual reading, respectively.

Afterward, the chips were divided into 5 groups. Four groups were exposed to 0.5, 1, 1.5, and 2 Gy, respectively, and the remaining group was used to measure the background dose. Batch calibration factor was then calculated by Eq. 2.



BCF is the batch calibration factor (Gy/nC) and CCdose(the correct count) is the TLD reading (nC). TLD dose (Gy) was calculated by Eq. 3 where BGDdose (Gy) is the background dose.



Finally, the chips were calibrated and read by an automatic double-channel reader (SOLARO 2A, NEC company) and annealed again for reuse.

Dose measurements

In the present study, ten TLD chips were evenly distributed throughout the left lung of the thorax phantom.

Accurate TLD placement is an important factor in determining the measured dose. A study by Herbert et al. showed that TLD positioning errors, caused by changes in the SSD, patient contour, and TLD response would lead to changes in the measured dose. To avoid TLD placement error, each set of TLD positions was repeated 3 times and the results were averaged.

The chips were placed in the phantom, and then, a CT scan was performed using an MDCT-64 (Siemens, SOMATOM Sensation). According to the RTOG breast contouring  Atlas More Details,[19] the target and organs at risk including lung, heart, and spinal cord were contoured by the oncologist.

The images were transferred to TPS (TiGRT, LinaTech, China). The usual RT technique in breast cancer with lymph node involvement is two tangential opposing fields and two anterior–posterior supraclavicular fields. The phantom was irradiated by 6 MV photons with a total dose of 50 Gy, using the linear accelerator Siemens Primus. Finally, the TPS dose calculations, derived from the dose–volume histograms, were compared with TLD dose measurements.

Statistical analysis

Paired sample t-test was used to compare mean TLD and TPS dose. Statistical analysis was performed with SPSS 22.0 (SPSS, Chicago, IL, USA). P value of 0.05 was considered statistically significant.


  Results Top


Mean TLD and TPS doses were 42.12 (SD = 13.22) and 43.26 (SD = 10.78), respectively. The TLD and TPS doses along with their mean differences, obtained by Eq. 4, are listed in [Table 1] as separate columns.
Table 1: Treatment planning system and thermoluminescent dosimeters dose in the lung

Click here to view




The results showed a significant difference between the mean doses (P = 0.007). The minimum and maximum difference were −0.91% and 4.46%, respectively. Higher dose differences were observed in the inner margin of the lung and breast tissue (TLD numbers 2, 3, 6, 7, and 8).




  Discussion Top


Lung exposure is unavoidable in breast or chest wall irradiation, with or without regional lymph node involvement. This incidental exposure may increase the risk of subsequent lung damage, such as pneumonitis.

In breast cancer RT, 3D-CRT is superior to IMRT as it uses less complex techniques, reduces the lung dose,[6],[7] and improves the conformity in the target volume.[20] Accurate lung dose assessment is particularly challenging because of the breast or chest wall deformities[8] and dose-dependent side effects.[21] The purpose of this study was to evaluate the lung dose in left-sided breast cancer with 3D-CRT by comparing TLD dose measurements in a thorax phantom with TPS dose calculations.

Castro et al.,[22] according to the detailed analysis of the International Atomic Energy Agency,[23] assessed a 5.8% TLD uncertainty for the megavoltage photon beams. The final standard uncertainty of 5.8% is because of repetitive TLD measurement uncertainties, TLD and Linac calibration uncertainties, absorbed dose energy dependence, TLD positioning uncertainties, and energy-dependence corrections. In another study, Almond et al.[24] determined a 5% TLD readout uncertainty by considering repetitive TLD measurements of 2.2%, TLD calibration of 1.8%, TLD positioning uncertainty of 0.2%, and energy dependence correction of 0.8%.

McCullough and Krueger[25] and Van Dyk et al.[26] suggested that the acceptable difference between TPS dose calculations and TLD dose measurements for external photon beams is 5%.

The results of the current study showed a 5% TPS dose-overestimation compared to TLD dose measurements. The dose difference might be due to limitations of the TPS dose calculation algorithm in inhomogeneous regions like the lung,[18] inaccurate beam modeling caused by contaminated electrons in low-density regions like lung,[27] and inherent limitations of the TLD.[28] Another result showed that the TLDs that were on the inner margin of the lung and breast tissue (TLD numbers 2, 3, 6, 7, and 8) showed higher dose differences than the TLDs in the lung. Dose differences are due to large density differences caused by a greater range of electrons in the lung.[29]

Our results were consistent with those of other studies. Baird et al.[30] compared 3D-TPS dose calculations with TLD dose measurements in the lung and concluded that the measured and calculated dose differed because of the limitations of the TPS dose calculation algorithm in regions of inhomogeneities. In a study by Farhood et al.,[27] TPS dose calculations in the build-up region of the tangential field of the breast were compared with TLD measurements. The results of their study showed that TPS overestimated doses compared to TLD measurements. They reached a conclusion that the dose overestimation may be due to inaccurate modeling of the dose contributions from contaminated electrons and secondary scatter photons derived from the accelerator head.

In addition, Zhao et al.[31] showed a 3% TPS dose-overestimation compared to TLD dose measurements in a water phantom. Butson et al.[14] used a male anthropomorphic phantom and placed 1.5 mm solid water between each slab. The solid water increased the complexity of dose measurement and verification and produced a distortion in the results. They showed a 5% TPS dose-overestimation in 3D-CRT. Other studies showed greater dose-overestimations. Davidson et al.[15] used an anthropomorphic phantom with polyvinylchloride plates, a nylon heart, a polybutadiene spine, and a proprietary material representing lung tissue and indicated a 10% dose-overestimation in IMRT treatment. Aljarrah et al.[32] performed a study with a prefabricated lung phantom and reported a 20% dose-overestimation in the IMRT treatment. As was expected, the overestimation was larger for the IMRT treatment compared to the 3D-CRT because of the higher dose received by the lung.

In the future work, TPS dose calculations could be compared by another dose calculation algorithm such as Monte Carlo and dose calculations could be studied in vivo in patients.




  Conclusions Top


The results of the current study showed up to 5% difference between dose measurements and calculations. TPS dose-overestimations were due to incorrect beam modeling caused by contaminated electrons in low-density regions like lung. Moreover, greater dose differences in the inner margin of the lung and breast tissue were as a result of large density variations.

Acknowledgments

We sincerely thank the staff of Isfahan Omid hospital for their help and support during this project and A. Hassanzadeh who assisted in processing the data.

Financial support and sponsorship

This study was funded by the Isfahan University of Medical Sciences (grant number 396959).

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Overgaard M, Christensen JJ, Johansen H, Nybo-Rasmussen A, Brincker H, van der Kooy P, et al. Postmastectomy irradiation in high-risk breast cancer patients. Present status of the Danish Breast Cancer Cooperative Group trials. Acta Oncol 1988;27:707-14.  Back to cited text no. 1
    
2.
Bellon JR, Katz A, Taghian A. Radiation therapy for breast cancer. Hematol Oncol Clin North Am 2006;20:239-57, vii.  Back to cited text no. 2
    
3.
Darby SC, McGale P, Taylor CW, Peto R. Long-term mortality from heart disease and lung cancer after radiotherapy for early breast cancer: Prospective cohort study of about 300,000 women in US SEER cancer registries. Lancet Oncol 2005;6:557-65.  Back to cited text no. 3
    
4.
Freedman GM, Anderson PR, Li J, Eisenberg DF, Hanlon AL, Wang L, et al. Intensity modulated radiation therapy (IMRT) decreases acute skin toxicity for women receiving radiation for breast cancer. Am J Clin Oncol 2006;29:66-70.  Back to cited text no. 4
    
5.
Senthilkumar S. Design of homogeneous and heterogeneous human equivalent thorax phantom for tissue inhomogeneity dose correction using TLD and TPS measurements. Int J Radiat Res 2014;12:179.  Back to cited text no. 5
    
6.
Xie X, Ouyang S, Wang H, Yang W, Jin H, Hu B, et al. Dosimetric comparison of left-sided whole breast irradiation with 3D-CRT, IP-IMRT and hybrid IMRT. Oncol Rep 2014;31:2195-205.  Back to cited text no. 6
    
7.
Prior P, Sparks I, Wilson J, Li X, White J. Use of three dimensional conformal radiation therapy (3DCRT) for node positive breast cancer does not result in excess lung and heart irradiation. Int J Rad Oncolo Biol Phys 2011;81:S805.  Back to cited text no. 7
    
8.
Mehta VK, Goffinet DR. Unsuspected abnormalities noted on CT treatment-planning scans obtained for breast and chest wall irradiation. Int J Radiat Oncol Biol Phys 2001;49:723-5.  Back to cited text no. 8
    
9.
Aznar MC, Duane FK, Darby SC, Wang Z, Taylor CW. Exposure of the lungs in breast cancer radiotherapy: A systematic review of lung doses published 2010-2015. Radiother Oncol 2018;126:148-54.  Back to cited text no. 9
    
10.
Borzone G, Moreno R, Urrea R, Meneses M, Oyarzún M, Lisboa C. Bleomycin-induced chronic lung damage does not resemble human idiopathic pulmonary fibrosis. Am J Respir Crit Care Med 2001;163:1648-53.  Back to cited text no. 10
    
11.
Lind PA, Gagliardi G, Wennberg B, Fornander T. A descriptive study of pulmonary complications after postoperative radiation therapy in node-positive stage II breast cancer. Acta Oncol 1997;36:509-15.  Back to cited text no. 11
    
12.
Herbert CE, Ebert MA, Joseph DJ. Feasible measurement errors when undertaking in vivo dosimetry during external beam radiotherapy of the breast. Med Dosim 2003;28:45-8.  Back to cited text no. 12
    
13.
Wong S, Back M, Tan PW, Lee KM, Baggarley S, Lu JJ. Can radiation therapy treatment planning system accurately predict surface doses in postmastectomy radiation therapy patients? Med Dosim 2012;37:163-9.  Back to cited text no. 13
    
14.
Butson MJ, Elferink R, Cheung T, Yu PK, Stokes M, Quach KY, et al. Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution method. Phys Med Biol 2000;45:N143-9.  Back to cited text no. 14
    
15.
Davidson SE, Ibbott GS, Prado KL, Dong L, Liao Z, Followill DS. Accuracy of two heterogeneity dose calculation algorithms for IMRT in treatment plans designed using an anthropomorphic thorax phantom. Med Phys 2007;34:1850-7.  Back to cited text no. 15
    
16.
Goldstone K. Tissue substitutes in radiation dosimetry and measurement. In: ICRU Report 44, International Commission on Radiation Units and Measurements. USA: W.B. Saunders; 1990.  Back to cited text no. 16
    
17.
Kelly CA, Wang XY, Chu JC, Hartsell WF. Dose to contralateral breast: A comparison of four primary breast irradiation techniques. Int J Radiat Oncol Biol Phys 1996;34:727-32.  Back to cited text no. 17
    
18.
Corredor CE. Dose Analysis by Radiation Treatment Planning System (TPS) Software vs. Thermoluminescent Dosimeters Output; 2004.  Back to cited text no. 18
    
19.
White J, Tai A, Arthur D, Buchholz T, MacDonald S, Marks L, et al. Breast cancer atlas for radiation therapy planning: Consensus definitions. Book Breast Cancer Atlas for Radiation Therapy Planning. 2009;73:944-51.  Back to cited text no. 19
    
20.
Haciislamoglu E, Colak F, Canyilmaz E, Zengin AY, Yilmaz AH, Yoney A, et al. The choice of multi-beam IMRT for whole breast radiotherapy in early-stage right breast cancer. Springerplus 2016;5:688.  Back to cited text no. 20
    
21.
Mauch P, Ng A, Aleman B, Carde P, Constine L, Diehl V, et al. Report from the rockefellar foundation sponsored international workshop on reducing mortality and improving quality of life in long-term survivors of Hodgkin's disease: July 9-16, 2003, Bellagio, Italy. Eur J Haematol Suppl 2005;66:68-76.  Back to cited text no. 21
    
22.
Castro P, García-Vicente F, Mínguez C, Floriano A, Sevillano D, Pérez L, et al. Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration. J Appl Clin Med Phys 2008;9:2676.  Back to cited text no. 22
    
23.
International Atomic Energy Agency. 398. Absorbed dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed dose to Water. Vienna: International Atomic Energy Agency; 2000.  Back to cited text no. 23
    
24.
Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, et al. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26:1847-70.  Back to cited text no. 24
    
25.
McCullough EC, Krueger AM. Performance evaluation of computerized treatment planning systems for radiotherapy: External photon beams. Int J Radiat Oncol Biol Phys 1980;6:1599-605.  Back to cited text no. 25
    
26.
Van Dyk J, Barnett RB, Cygler JE, Shragge PC. Commissioning and quality assurance of treatment planning computers. Int J Radiat Oncol Biol Phys 1993;26:261-73.  Back to cited text no. 26
    
27.
Farhood B, Bahreyni Toossi MT, Soleymanifard S, Mortezazadeh T. Assessment of the accuracy of dose calculation in the build-up region of the tangential field of the breast for a radiotherapy treatment planning system. Contemp Oncol (Pozn) 2017;21:232-9.  Back to cited text no. 27
    
28.
Nikoofar A, Hoseinpour Z, Mahdavi SR, Hasanzadeh H, Tavirani MR. High-dose-rate 192Ir brachytherapy dose verification: A phantom study. Iran J Cancer Prev 2015;8:23-31.  Back to cited text no. 28
    
29.
Ottosson W, Andersen CE, Behrens CF. Improved radiotherapy for locally advanced non-small cell lung carcinoma (NSCLC) patients 2015;12:35-43.  Back to cited text no. 29
    
30.
Baird CT, Starkschall G, Liu HH, Buchholz TA, Hogstrom KR. Verification of tangential breast treatment dose calculations in a commercial 3D treatment planning system. J Appl Clin Med Phys 2001;2:73-84.  Back to cited text no. 30
    
31.
Zhao Y, Qi G, Yin G, Wang X, Wang P, Li J, et al. A clinical study of lung cancer dose calculation accuracy with Monte Carlo simulation. Radiat Oncol 2014;9:287.  Back to cited text no. 31
    
32.
Aljarrah K, Pawlickj T, Niemierko A. A clinical study of MLC-based IMTT lung dose calculation accuracy on plan evaluation parameters. J Cancer Sci Ther 2010;2:74.  Back to cited text no. 32
    


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
   Abstract
  Introduction
  Methods
  Results
  Discussion
  Conclusions
   References
   Article Figures
   Article Tables

 Article Access Statistics
    Viewed69    
    Printed0    
    Emailed0    
    PDF Downloaded22    
    Comments [Add]    

Recommend this journal