Introduction of a new parameter for evaluation of digital radiography system performance


1 Student Research Committee, Faculty of Paramedicine, Isfahan University of Medical Sciences, Isfahan, Iran
2 Technology of Radiology Department, Isfahan University of Medical Sciences, Isfahan, Iran

Correspondence Address:
Prof. Ali Chaparian
Department of Technology of Radiology, Isfahan University of Medical Sciences, Isfahan
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmss.JMSS_53_19

Rights and Permissions

Background: The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom. Methods: The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD phantom. Incident air kerma (IAK) values were measured for certain exposure settings in all digital radiography systems. The images from the CDRAD phantom were evaluated by three observers. The results were displayed in the form of a contrast-detail (CD) curve. In addition, the inverse image quality figure (IQFinv)-to-IAK ratios were used for quantitative comparison of different digital radiography system performance. Results: Results of this study showed that the CD curves cannot be suitable criterion for determining the performance of digital radiography systems. For this reason, IQFinv-to-radiation dose (IAK) ratios in a fixed radiation condition were used. The highest performance in terms of producing high-quality images and low radiation dose was related to X-ray unit 1 and the lowest performance was for X-ray unit 5. Conclusion: The ratio of IQFinv to IAK for performance evaluation of digital radiography systems is an innovation of this study. A digital radiography system with a higher IQFinv-to-IAK ratio is associated with lower patient dose and better image quality. Therefore, it is recommended to equip the new imaging centers with the systems that have higher IQFinv-to-IAK ratios.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed260    
    Printed19    
    Emailed0    
    PDF Downloaded47    
    Comments [Add]    

Recommend this journal